QRS - PC-Based Universal ECG Machine
The QRS PC Based ECG is an easy-to-use, small and light ECG machine. Store and analyze up to 12-channels of high-resolution ECG data. The QRS ECG interfaces with Office Medic, which is able to integrate with Best Practice & Medical Director.
QRS PC-Based Universal ECG Features
Why Choose the QRS PC-Based Universal ECG
Specifications of the QRS PC-Based Universal ECG
Hub Weight | 250-335 grams (depending on cables) |
Hub Dimensions | 85 mm x 91 mm x 20 mm (3.3” x 3.6” x .8”) |
Patient Leads Length | Limb Leads: 1 meter Chest Leads: 0.6 meter |
PC Connection Length | 1-3 meters USB A type Connector (USB version) |
Patient Leads | 12-Lead Cable (10 Patient Leads) |
Case Material | ABS Plastic |
Electrode Connection | 4 mm Banana plug with “snap” or “tab” connectors |
Display & Operating Console | Dependent on PC (Supplied by user) |
Gain Sensitivity | 5, 10, 20 mm/mV |
Speed | 12.5, 25 and 50 mm/s |
Input Range | ±6 mV |
Acquisition Sample Rate | 1000 samples/sec.
(Compressed to 500 Hz with peak picking and averaging algorithm) |
Heart Rate Range | 30 – 170 bpm |
Frequency Response | 0.05 – 175 Hz ±3dB |
Defibrillation Protection | Patient leads are isolated from system and operator, with 4kV protection |
Common Mode Rejection | -60dB (minimum) |
Safety Standards | Complies with AAMI EC11, EN60601-1, EN60601-2, EN60601-2-25 |
Leads Off Indicators | Connection status for each lead is shown on Acquisition screen |
Power Source | PC, powered from USB interface |
Supply Voltage | 4-16 V DC |
Supply Current | < 17 mA DC |
Permanent Filters | High Pass: 0.05 Hz 1st order; Low Pass: 170 Hz 1st order
Baseline Wander: Baseline reset by adaptive zeroing |
Notch filter
(Mains noise Rejection) |
50 Hz 4th order Butterworth,
49.1 Hz – 50.9 Hz, 60 Hz 4th order Butterworth, 59.1 Hz – 60.9 Hz |
Low Pass
(Muscle Artifact Filter) |
35 Hz 4th order |
Report Capabilities | User Selectable Report Formats: (1) 3×4 Simultaneous; (2) 3×4 Sequential; (3) 6×2 ECG; (4) 6×1 ECG; (5) Averaged Complexes ECG; (6) ECG Measurements Report; (7) Avg. Zoom Report; (8) Zoom Report |
Interpretation | Rhythm, Morphology |
Environmental Conditions | Operating Temperature: 0-40°C
Storage Conditions: -20-70°C Humidity: 85% Max. (non-condensing) |
Contact Supplier for more information
|