BioSan Cell Counter | Densitometer Suspension ...
Key Features
Product Description
The DEN-1B densitometer from BioSan has been developed to measure the turbidity of a solution containing cells or microorganisms. Turbidity is measured over a range of 0-15 McFarland units.
Key applications for the DEN-1B include:
The DEN-1B is compatible with glass and transparent plastic 12 mm and 16 mm tubes (with supplied A-12 and A-16 adapters) or 18 mm tubes without the adapter. The instrument is factory-calibrated for use with 16 mm diameter glass tubes with calibration parameters being stored in the DEN-1B in case the unit needs to be reset to factory calibration settings.
The DEN-1B should, however, be recalibrated to the diameter and configuration of the end-user's sample tubes if these are not 16 mm glass tubes. It should also be recalibrated any time another type of sample tube is used. Optional turbidity standards are available (please see below) or the end-user can employ their own standards for this purpose.
Optional Accessories
The following tubes and calibration kits are available for the DEN-1B:
Part Number | Item Description |
CKG16 | Set of 0.5, 1.0, 2.0, 3.0 and 4.0 McFarland Turbidity Standards (latex particles): for 16 mm diameter glass tubes |
CKG1802 | Set of 0.5, 1.0, 2.0, 3.0, 4.0 and 5.0 McFarland Turbidity Standards (BaSO4): for 18 mm diameter glass tubes |
CKG12 | Set of 0.0 (blank), 0.5, 2.0 and 3.0 McFarland Turbidity Standards (latex particles): for 12 mm diameter glass tubes |
BS-050102-LK | Glass sample tubes without cap, (16 mm ? 100 mm x 0.8 mm), quantity of 78 tubes |
Specifications
Manufacturer | Biosan |
Model number | DEN-1B |
Measurement range | 0-15 McFarland units |
Display resolution | 0.01 McFarland units |
Light source | LED |
Measurement wavelength (λ) | 565 nm ±15 nm |
Accuracy (0 - 6 McF) | ±3% |
Measurement time | 1 second |
Sample volume | ≥2 mL |
Tube external diameter | 12 mm, 16 mm (using A-12 and A-16 adapters) or 18 mm (without adapter) |
Restore factory calibration settings | Yes |
Display | LCD |
Overall dimensions (W x D x H) | 165 mm x 115 mm x 75 mm |
Weight | 0.7 Kg |
Input current | power consumption | 12 V | 7 mA: 0.1 W |
External power supply (supplied with instrument) | Input: 100–240 VAC, 50/60 Hz | Output: 12 VDC |
Independent power supply | 3 x AA batteries |
To contact Advance Scientific Products Pty Ltd about BioSan Cell Counter | Densitometer Suspension Turbidity Quantification- DEN1B use Get a quote.
Key Features
Product Description
The DEN-1B densitometer from BioSan has been developed to measure the turbidity of a solution containing cells or microorganisms. Turbidity is measured over a range of 0-15 McFarland units.
Key applications for the DEN-1B include:
The DEN-1B is compatible with glass and transparent plastic 12 mm and 16 mm tubes (with supplied A-12 and A-16 adapters) or 18 mm tubes without the adapter. The instrument is factory-calibrated for use with 16 mm diameter glass tubes with calibration parameters being stored in the DEN-1B in case the unit needs to be reset to factory calibration settings.
The DEN-1B should, however, be recalibrated to the diameter and configuration of the end-user's sample tubes if these are not 16 mm glass tubes. It should also be recalibrated any time another type of sample tube is used. Optional turbidity standards are available (please see below) or the end-user can employ their own standards for this purpose.
Optional Accessories
The following tubes and calibration kits are available for the DEN-1B:
Part Number | Item Description |
CKG16 | Set of 0.5, 1.0, 2.0, 3.0 and 4.0 McFarland Turbidity Standards (latex particles): for 16 mm diameter glass tubes |
CKG1802 | Set of 0.5, 1.0, 2.0, 3.0, 4.0 and 5.0 McFarland Turbidity Standards (BaSO4): for 18 mm diameter glass tubes |
CKG12 | Set of 0.0 (blank), 0.5, 2.0 and 3.0 McFarland Turbidity Standards (latex particles): for 12 mm diameter glass tubes |
BS-050102-LK | Glass sample tubes without cap, (16 mm ? 100 mm x 0.8 mm), quantity of 78 tubes |
Specifications
Manufacturer | Biosan |
Model number | DEN-1B |
Measurement range | 0-15 McFarland units |
Display resolution | 0.01 McFarland units |
Light source | LED |
Measurement wavelength (λ) | 565 nm ±15 nm |
Accuracy (0 - 6 McF) | ±3% |
Measurement time | 1 second |
Sample volume | ≥2 mL |
Tube external diameter | 12 mm, 16 mm (using A-12 and A-16 adapters) or 18 mm (without adapter) |
Restore factory calibration settings | Yes |
Display | LCD |
Overall dimensions (W x D x H) | 165 mm x 115 mm x 75 mm |
Weight | 0.7 Kg |
Input current | power consumption | 12 V | 7 mA: 0.1 W |
External power supply (supplied with instrument) | Input: 100–240 VAC, 50/60 Hz | Output: 12 VDC |
Independent power supply | 3 x AA batteries |